The Nuvair Pro CO2 panel mount alarm analyzers comes in three power options, two module options, and two sensor options. The CO2 sensor is auto-calibrating and provides a temperature-compensated, linear CO2 measurement over sensing range.
Two audible and visual alarms for low and high carbon dioxide content can be set by the user. If the alarm sounds, the analog output can activate an optional relay that shuts down compressors automatically, sounds remote alarms, or activates emergency backup systems.
WARNING: Never expose gas sensors to pressure or you may cause damage and/or false readings. Damaged sensors will not provide accurate gas analysis. Most gas analyzers can be used to analyze a regulated gas sample flow, the contents of a gas cylinder, or the flow from a regulator. The flow rate of gas must equal 1-5 L/min. To produce this flow, a Flow Restrictor and Regulator may be required. A faulty Flow Restrictor can lead to a false analyzer reading. Flow Restrictors should be regularly tested with a Flow Meter. Inaccurate gas analysis can lead to serious personal injury or death.
Technical Specifications
Product Code: SY-NUVAIR-SKU-9617
• Flow Rate 0.6–1 L/min
• Range 0–2000 ppm (Optional 0–5000 ppm)
• Resolution 50 ppm (0–1000 ppm), then 100 ppm up to full scale
• Optional Resolution 5000 ppm with Lithium-Ion Battery & Charger
• Repeatability ±2% of Full Scale @ 68°F (20°C)
• Linearity ±2% FSD or ±10% of reading (whichever is greater)
• Sensor Type NDIR (Non-Dispersive Infrared)
• Expected Sensor Life 5–10 Years (Subject to Use & Maintenance)
• Alarms Two User-Programmable Audible & Visual Alarms
• Response Time < 30 Seconds @ 68°F (20°C)
• Operating & Storage Temperature –4°F to +122°F (–20°C to +50°C)
• Operating Humidity 0–95% RH, Non-Condensing
• Power Options 110/230V Plug-In / Rechargeable Lithium Battery / DIN Rail (9617 Only)
• Battery Life Approx. 20 Hours Per Full Charge
• Dimensions 4 x 2 x 5.5 in (10.2 x 5 x 14 cm)
• Weight 9 oz (255 g)
• Warranty 12 Months